MycoFog ® FAQs & Troubleshooting Guide
1. What is MycoFog®?
MycoFog® is an advanced biodecontamination system that uses hydrogen peroxide (H2O2) fogging to eliminate microbial contamination from incubators, workstations, and other enclosed spaces.
2. What reagents can I use with MycoFog®?
- MycoFog® H2O2 reagents (our approved reagent for optimal performance):
| MycoFog® reagent, 4 bottles + 4 wicks. For incubators <200L | MFR-1Bx-K |
| MycoFog® reagent, 4 bottles + 4 wicks. For incubators >200L and <350L | MFR-2Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >350L and <500L | MFR-3Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >500L and <700L | MFR-4Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >700L and <1000L | MFR-5Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >1000L and <1200L | MFR-6Bx-K |
- Other compatible H2O2 solutions can be used (but please ensure proper concentration and purity).
Check out the link for MycoFog® reagents
3. How often should I perform a MycoFog® decontamination cycle?
For optimal contamination control, we recommend performing a MycoFog® cycle once per month per incubator.
4. How do I prepare MycoFog® for a decontamination cycle?
Ensure your incubator is clean! The MycoFog® biodecontamination process complements cleaning but does not replace it. Please consult your incubator operating manual for the manufacturer’s cleaning protocol.
Ensure the incubator is empty (remove cultures, labware, and water pans).
Set the temperature to 45°C (or use the 37°C protocol for 4 hours).
Disable gas control (CO2 and humidity settings).
Place MycoFog® at the mid-level inside the incubator for optimal fog distribution.
Press the function button to start the cycle.
5. How long does the MycoFog® cycle take?
Standard Cycle: 180 minutes at 45°C
37°C Protocol: 4 hours for <200L chambers, 5 hours for larger chambers
6. Can I use MycoFog® in large incubators?
Yes! There are 2 instrument models of MycoFog®:
– Model MF-1B for incubators up to <500L internal volume
– Model MF-500 for incubators >500L but <1200L internal volume
MycoFog® reagents come in 6 different pack sizes, each formulation is designed for a size of Incubator, Glovebox, or Workstation.
Take care to order the correct MycoFog® Reagent pack for your instrument.
Table: Choose the right MycoFog® products
| Table to help Customers choose the right MycoFog® products | ||||
|---|---|---|---|---|
| MycoFog® | Incubator or Workstation to be decontaminated - Internal Vol (L) | MycoFog® Instrument needed | MycoFog® Reagent Kit Product code needed | MycoFog® Starter Kit needed |
| <200L | MF 1B | MF-1Bx | MF-SK-1Bx | |
| >200L <350L | MF-2Bx | MF-SK-2Bx | ||
| >350L <500L | MF-2Bx | MF-SK-3Bx | ||
| >500L <700L | MF-500 | MFR-4Bx-K | MF500-SK-4Bx | |
| >700L <1000L | MFR-5Bx-K | MF500-SK-5Bx | ||
| >1000L <1200L | MFR-6Bx-K | MF500-SK-6Bx | ||
7. What is the recommended relative humidity before starting a cycle?
MycoFog® works best when the incubator’s starting relative humidity is ambient.
The device achieves proper humidity during the cycle.
8. Can I use MycoFog® in an incubator that contains powered electronics?
Yes. MycoFog® has been tested with electronic devices such as the Etaluma Lumascope and the Sartorius Incucyte, which perform normally after repeated decontamination cycles.
9. My MycoFog® instrument is not producing enough spray. What should I do?
Check the Wick: Remove wick, ensure it is fully saturated, and reseat wick. Ensure wick holder is seated properly.
Check the Reagent: Ensure the correct amount is added and that only MycoFog® H2O2 reagent or an approved formulation is used.
Clean the Fogger: Use a cotton swab with vinegar to clean the nebulizer element.
Charge the Battery: Low power can reduce spray intensity — fully charge before use.
10. My MycoFog® stopped mid-cycle. How do I fix it?
Battery may be low — recharge the unit.
Restart the device by turning it off and on again.
11. The reagent is leaking — what should I check?
Ensure the reagent reservoir is properly sealed.
Confirm that the instrument is level and not tilted.
Do not overfill — use only the recommended reagent volume.
12. How long does the battery last?
A fully charged battery will perform approximately 10 cycles.
We recommend fully charging the instrument before each use if repeated uses are done on different days.
13. Can MycoFog® operate while charging?
Yes, MycoFog® can function while connected to a power source.
14. How do I clean my MycoFog® instrument?
After each use, rinse the reagent reservoir with distilled water.
If fogging volume decreases, clean the nebulizer disk with vinegar and a cotton swab.
Use a fresh wick for each cycle.
15. What quality control checks are performed before shipping?
Each MycoFog® instrument is tested to ensure:
– Proper flow rate within the acceptable range
– Residual reagent volume is below the maximum allowable level
16. Can MycoFog® be used in a GMP environment?
Yes! Our solvent-resistant instrument labels allow for alcohol wipe-downs before GMP suite entry.
17. How do I validate MycoFog®’s effectiveness?
Use an approved Biological Indicator (BI’s) to confirm a reduction in microbial contamination.
18. Where can I purchase MycoFog® and MycoFog® reagents?
For US/Canada, visit our website at https://mycofog.com and click “Get MycoFog®”.
For international customers, please visit https://mycofog.com and contact your authorized Partner.
19. Need Help?
For troubleshooting, orders, or technical support, contact us at:
[email protected]
MycoFog ® FAQs & Troubleshooting Guide
1. What is MycoFog®?
2. What reagents can I use with MycoFog®?
- MycoFog® H2O2 reagents (our approved reagent for optimal performance):
| MycoFog® reagent, 4 bottles + 4 wicks. For incubators <200L | MFR-1Bx-K |
| MycoFog® reagent, 4 bottles + 4 wicks. For incubators >200L and <350L | MFR-2Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >350L and <500L | MFR-3Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >500L and <700L | MFR-4Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >700L and <1000L | MFR-5Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >1000L and <1200L | MFR-6Bx-K |
- Other compatible H2O2 solutions can be used (but please ensure proper concentration and purity).
Check out the link for MycoFog® reagents
3. How often should I perform a MycoFog® decontamination cycle?
Ensure your incubator is clean! The MycoFog® biodecontamination process complements cleaning but does not replace it. Please consult your incubator operating manual for the manufacturer’s cleaning protocol.
Ensure the incubator is empty (remove cultures, labware, and water pans).
Set the temperature to 45°C (or use the 37°C protocol for 4 hours).
Disable gas control (CO2 and humidity settings).
Place MycoFog® at the mid-level inside the incubator for optimal fog distribution.
Press the function button to start the cycle.
5. How long does the MycoFog® cycle take?
Standard Cycle: 180 minutes at 45°C
37°C Protocol: 4 hours for <200L chambers, 5 hours for larger chambers
6. Can I use MycoFog® in large incubators?
Yes! There are 2 instrument models of MycoFog®:
– Model MF-1B for incubators up to <500L internal volume
– Model MF-500 for incubators >500L but <1200L internal volume
MycoFog® reagents come in 6 different pack sizes, each formulation is designed for a size of Incubator, Glovebox, or Workstation.
Take care to order the correct MycoFog® Reagent pack for your instrument.
| Table to help Customers choose the right MycoFog® products | ||||
|---|---|---|---|---|
| MycoFog® | Incubator or Workstation to be decontaminated – Internal Vol (L) | MycoFog® Instrument needed | MycoFog® Reagent Kit Product code needed | MycoFog® Starter Kit needed |
| <200L | MF 1B | MF-1Bx | MF-SK-1Bx | |
| >200L <350L | MF-2Bx | MF-SK-2Bx | ||
| >350L <500L | MF-2Bx | MF-SK-3Bx | ||
| >500L <700L | MF-500 | MFR-4Bx-K | MF500-SK-4Bx | |
| >700L <1000L | MFR-5Bx-K | MF500-SK-5Bx | ||
| >1000L <1200L | MFR-6Bx-K | MF500-SK-6Bx | ||
MycoFog® works best when the incubator’s starting relative humidity is ambient.
The device achieves proper humidity during the cycle.
8. Can I use MycoFog® in an incubator that contains powered electronics?
Yes. MycoFog® has been tested with electronic devices such as the Etaluma Lumascope and the Sartorius Incucyte, which perform normally after repeated decontamination cycles.
Check the Wick: Remove wick, ensure it is fully saturated, and reseat wick. Ensure wick holder is seated properly.
Check the Reagent: Ensure the correct amount is added and that only MycoFog® H2O2 reagent or an approved formulation is used.
Clean the Fogger: Use a cotton swab with vinegar to clean the nebulizer element.
Charge the Battery: Low power can reduce spray intensity — fully charge before use.
10. My MycoFog® stopped mid-cycle. How do I fix it?
Battery may be low — recharge the unit.
Restart the device by turning it off and on again.
11. The reagent is leaking — what should I check?
Ensure the reagent reservoir is properly sealed.
Confirm that the instrument is level and not tilted.
Do not overfill — use only the recommended reagent volume.
A fully charged battery will perform approximately 10 cycles.
We recommend fully charging the instrument before each use if repeated uses are done on different days.
13. Can MycoFog® operate while charging?
Yes, MycoFog® can function while connected to a power source.
14. How do I clean my MycoFog® instrument?
After each use, rinse the reagent reservoir with distilled water.
If fogging volume decreases, clean the nebulizer disk with vinegar and a cotton swab.
Use a fresh wick for each cycle.
15. What quality control checks are performed before shipping?
Each MycoFog® instrument is tested to ensure:
– Proper flow rate within the acceptable range
– Residual reagent volume is below the maximum allowable level
Yes! Our solvent-resistant instrument labels allow for alcohol wipe-downs before GMP suite entry.
17. How do I validate MycoFog®’s effectiveness?
Use an approved Biological Indicator (BI’s) to confirm a reduction in microbial contamination.
For US/Canada, visit our website at https://mycofog.com and click “Get MycoFog®”.
For international customers, please visit https://mycofog.com and contact your authorized Partner.
19. Need Help?
For troubleshooting, orders, or technical support, contact us at:
[email protected]
MycoFog ® FAQs & Troubleshooting Guide
MycoFog® is an advanced biodecontamination system that uses hydrogen peroxide (H2O2) fogging to eliminate microbial contamination from incubators, workstations, and other enclosed spaces.
2. What reagents can I use with MycoFog®?
- MycoFog® H2O2 reagents (our approved reagent for optimal performance):
| MycoFog® reagent, 4 bottles + 4 wicks. For incubators <200L | MFR-1Bx-K |
| MycoFog® reagent, 4 bottles + 4 wicks. For incubators >200L and <350L | MFR-2Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >350L and <500L | MFR-3Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >500L and <700L | MFR-4Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >700L and <1000L | MFR-5Bx-K |
| MycoFog® reagent, 2 bottles + 2 wicks. For incubators >1000L and <1200L | MFR-6Bx-K |
- Other compatible H2O2 solutions can be used (but please ensure proper concentration and purity).
Check out the link for MycoFog® reagents
3. How often should I perform a MycoFog® decontamination cycle?
For optimal contamination control, we recommend performing a MycoFog® cycle once per month per incubator.
Ensure your incubator is clean! The MycoFog® biodecontamination process complements cleaning but does not replace it. Please consult your incubator operating manual for the manufacturer’s cleaning protocol.
Ensure the incubator is empty (remove cultures, labware, and water pans).
Set the temperature to 45°C (or use the 37°C protocol for 4 hours).
Disable gas control (CO2 and humidity settings).
Place MycoFog® at the mid-level inside the incubator for optimal fog distribution.
Press the function button to start the cycle.
5. How long does the MycoFog® cycle take?
Standard Cycle: 180 minutes at 45°C
37°C Protocol: 4 hours for <200L chambers, 5 hours for larger chambers
6. Can I use MycoFog® in large incubators?
Yes! There are 2 instrument models of MycoFog®:
– Model MF-1B for incubators up to <500L internal volume
– Model MF-500 for incubators >500L but <1200L internal volume
MycoFog® reagents come in 6 different pack sizes, each formulation is designed for a size of Incubator, Glovebox, or Workstation.
Take care to order the correct MycoFog® Reagent pack for your instrument.
| Table to help Customers choose the right MycoFog® products | ||||
|---|---|---|---|---|
| MycoFog® | Incubator or Workstation to be decontaminated – Internal Vol (L) | MycoFog® Instrument needed | MycoFog® Reagent Kit Product code needed | MycoFog® Starter Kit needed |
| <200L | MF 1B | MF-1Bx | MF-SK-1Bx | |
| >200L <350L | MF-2Bx | MF-SK-2Bx | ||
| >350L <500L | MF-2Bx | MF-SK-3Bx | ||
| >500L <700L | MF-500 | MFR-4Bx-K | MF500-SK-4Bx | |
| >700L <1000L | MFR-5Bx-K | MF500-SK-5Bx | ||
| >1000L <1200L | MFR-6Bx-K | MF500-SK-6Bx | ||
MycoFog® works best when the incubator’s starting relative humidity is ambient.
The device achieves proper humidity during the cycle.
8. Can I use MycoFog® in an incubator that contains powered electronics?
Yes. MycoFog® has been tested with electronic devices such as the Etaluma Lumascope and the Sartorius Incucyte, which perform normally after repeated decontamination cycles.
Check the Wick: Remove wick, ensure it is fully saturated, and reseat wick. Ensure wick holder is seated properly.
Check the Reagent: Ensure the correct amount is added and that only MycoFog® H2O2 reagent or an approved formulation is used.
Clean the Fogger: Use a cotton swab with vinegar to clean the nebulizer element.
Charge the Battery: Low power can reduce spray intensity — fully charge before use.
10. My MycoFog® stopped mid-cycle. How do I fix it?
Battery may be low — recharge the unit.
Restart the device by turning it off and on again.
11. The reagent is leaking — what should I check?
Ensure the reagent reservoir is properly sealed.
Confirm that the instrument is level and not tilted.
Do not overfill — use only the recommended reagent volume.
A fully charged battery will perform approximately 10 cycles.
We recommend fully charging the instrument before each use if repeated uses are done on different days.
13. Can MycoFog® operate while charging?
Yes, MycoFog® can function while connected to a power source.
14. How do I clean my MycoFog® instrument?
After each use, rinse the reagent reservoir with distilled water.
If fogging volume decreases, clean the nebulizer disk with vinegar and a cotton swab.
Use a fresh wick for each cycle.
15. What quality control checks are performed before shipping?
Each MycoFog® instrument is tested to ensure:
– Proper flow rate within the acceptable range
– Residual reagent volume is below the maximum allowable level
Yes! Our solvent-resistant instrument labels allow for alcohol wipe-downs before GMP suite entry.
17. How do I validate MycoFog®’s effectiveness?
Use an approved Biological Indicator (BI’s) to confirm a reduction in microbial contamination.
For US/Canada, visit our website at https://mycofog.com and click “Get MycoFog®”.
For international customers, please visit https://mycofog.com and contact your authorized Partner.
19. Need Help?
For troubleshooting, orders, or technical support, contact us at:
[email protected]
validation
7.8 BIO tests the MycoFog® biodecontamination process by placing the biological indicator (BI) (from Mesa Labs or Terragene) which are stainless steel discs impregnated with at least 1 x 106 Geobacillus stearothermophilus (G.s.) spores, into the incubator. The device is sealed so that it is required that the sterilant (nebulized hydrogen peroxide, nHP) penetrate a membrane to come into contact with the spores, killing them.


